Historic Advancement in Medical Accessibility: DOJ Adopts New Standards for Diagnostic Equipment

Washington, D.C. - In a move set to revolutionize healthcare for millions of Americans with disabilities, the Department of Justice (DOJ) has adopted new accessibility standards for medical diagnostic equipment (MDE). This landmark decision, announced on August 9, marks a milestone in the fight for equality in healthcare.

The new standards, which have already come into effect, will require public hospitals and clinics to adapt their diagnostic equipment, from examination tables to mammography machines, to make them accessible to people with various types of disabilities.


Experts estimate that this measure could significantly improve health outcomes for more than 61 million Americans with disabilities by eliminating barriers that often result in delayed or inadequate diagnoses. Below, we analyze in detail the impact and technical requirements of this new regulation:


Importance of the ADA and the New Regulation The Americans with Disabilities Act (ADA) of 1990 was a crucial step in protecting the rights of people with disabilities, but it did not specifically address the accessibility of medical equipment. The new DOJ regulation closes this critical gap, establishing clear and enforceable standards for MDE accessibility under the ADA.


Areas Impacted by the Regulation


1. Medical Equipment Design: Manufacturers will need to redesign equipment to meet the new accessibility standards.


2. Equipment Acquisition: Hospitals and clinics will have to invest in new accessible equipment or modify existing ones.


3. Staff Training: Additional training will be required for the proper use of new accessible equipment.


4. Patient Experience: Will significantly improve for those with disabilities, allowing easier and more dignified access to diagnostic services.


Detailed Technical Requirements


1. Transfer Height:

- Low height range: 17-19 inches (430-485 mm) from the floor

- High height: 25 inches (635 mm)

- Minimum of 4 intermediate positions separated by at least 1 inch (25 mm)


2. Transfer Surface Size:

- For end transfer: 28 inches (710 mm) minimum width, 17 inches (430 mm) minimum length

- For side transfer: 28 inches (710 mm) minimum width, 28 inches (710 mm) minimum length


3. Transfer Supports:

- Location: maximum 1.5 inches (38 mm) from the transfer surface

- Length: 15 inches (380 mm) minimum for end transfer, 28 inches (710 mm) minimum for side transfer

- Height: 6-19 inches (150-485 mm) above the transfer surface


4. Wheelchair Space:

- Width: 36 inches (915 mm) minimum

- Depth: varies by type of entry (48 inches/1220 mm for front or rear entry, 60 inches/1525 mm for side entry)


5. Mammography Platforms:

- Adjustability: from 26 inches (660 mm) to 42 inches (1065 mm) in height

- Knee clearance: 25 inches (635 mm) minimum depth, 28 inches (710 mm) maximum


6. Operable Controls:

- One-hand activation without tight grasping or wrist twisting

- Maximum activation force: 5 pounds (22.2 N)


Impact on Patients and Providers

For patients with disabilities, these changes mean safer and more dignified access to diagnostic services, potentially improving health outcomes. Medical providers, while facing initial implementation costs, could benefit in the long term from more efficient care and reduced risk of injury.


Conclusion: implementation and compliance

The success of this regulation will depend on its effective implementation. Entities covered by the ADA must ensure that their MDE complies with these standards or face potential legal action and penalties. It is crucial that healthcare providers take proactive measures to meet these new requirements.